Accurate phase calibration of PMUs and PMU calibrators

Proceedings of the 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016),
Ottawa, Canada, pp. 1 – 2 (2016).
Milos Acanski, Gert Rietveld, and Dennis Hoogenboom
.

Abstract
A measurement setup is developed for accurate phase calibration of phasor measurement units (PMUs) and PMU calibrators. The measurement method is based on generating a reference signal with known phase with respect to a 1 pulse-per-second time reference using a stepped sine wave signal. 

Click here for the publication